Z-2424-2010: NeoMarkers Rabbit Monoclonal anti-Human Cyclin D1 Antibody (Clone SP4), immunochemistry Reagent and Kit,

Recall record Z-2424-2010 for Lab Vision Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2424-2010
Company in FDA record
Lab Vision Corporation
FDA classification date
Sep 15, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Two lots may be contaminated with Ki-67 antibody, which could cause abnormal staining patterns, and produce false positive results.

Source facts

Action
Thermo Fisher Scientific issued an "Advisory Notice" letter dated March 5, 2010 to all domestic customers and international distributors. The letter described the product, problem and action to be taken. Customers were instructed to: 1) discard the affected lots, and 2) retest specimens using a replacement lot supplied by the firm, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 International distributors were instructed to: 1) place the affected lots and replace with a replacement lot supplied by the firm, and 2) provide customers with copies of the customer letter, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 For questions, contact Thermo Fisher Scientific at 1-800-828-1628 or 1-510-771-1595.
Initiated
2010-03-05
Root cause
Material/Component Contamination
Terminated
2011-03-21
Distribution
Worldwide Distribution: USA, including the states of CA, SC, IL NY, WI, PA, MN, AL, WA, MA, and KS, and the coutnries of Austria, Australia, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Malaysia, the Netherlands, Norway, Pakistan, Poland, Portugal, Russia, the UK, and Turkey.
Recall number
Z-2424-2010
Recall status
Terminated
CDRH record ID
93227
FDA classified
2010-09-15
Recall event number
56320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗