Open original FDA record ↗Recall record
- FDA identifier
- Z-2424-2018
- Company in FDA record
- Ion Beam Applications S.A.
- FDA classification date
- Jun 20, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
Source facts
- Action
- On May 18, 2018 IBA issued URGENT FIELD SAFETY NOTICES to their customers. Customers were advised to take the following actions: It is recommended to ensure that the correction vector confirmation message appears on the expected access point. When the correction vector is applied (if deemed necessary1), ensure that the Patient Positioning System GoTo the corrected position for treatment before starting irradiation: - verify on adaPTdeliver screen that the corrected target position is selected before engaging the movement. - verify on adaPTdeliver screen that the corrected target position appears on the screen after the movement. Moreover, as advised in the Proton Therapy System Clinical User Guide, it is recommended to verify the patient position after each correction vector application. A software update will be deployed on all affected sites by July 2018. Customers with questions may email: vigiliance@iba-group.com
- Initiated
- 2018-05-16
- Root cause
- Software design
- Terminated
- 2018-10-16
- Distribution
- Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
- Recall number
- Z-2424-2018
- Recall status
- Terminated
- CDRH record ID
- 165077
- FDA classified
- 2018-06-20
- Recall event number
- 80298