The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.
Source facts
Action
The firm notified its consignees by letter on 05/18/2020. The letter explained the issue, how it occurs, cautions and a workaround. The notices were disseminated to the following entities: Director of Biomedical/Clinical Engineering, Chief of Nursing, Health Care Administrator/Risk Manager.
Initiated
2020-05-18
Root cause
Software Design Change
Distribution
Worldwide Distribution
Recall number
Z-2424-2020
Recall status
Open, Classified
CDRH record ID
181744
FDA classified
2020-06-24
Recall event number
85765