Skip to content

Public FDA record

Z-2424-2020: CARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 800

Recall record Z-2424-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2424-2020
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jun 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.

Source facts

Action
The firm notified its consignees by letter on 05/18/2020. The letter explained the issue, how it occurs, cautions and a workaround. The notices were disseminated to the following entities: Director of Biomedical/Clinical Engineering, Chief of Nursing, Health Care Administrator/Risk Manager.
Initiated
2020-05-18
Root cause
Software Design Change
Distribution
Worldwide Distribution
Recall number
Z-2424-2020
Recall status
Open, Classified
CDRH record ID
181744
FDA classified
2020-06-24
Recall event number
85765

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗