Skip to content

Public FDA record

Z-2425-2018: PD-L1 IHC 22C3 pharmDx is a companion diagnostic used to determine PD-L1 levels for determination of pati

Recall record Z-2425-2018 for Dako North America Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2425-2018
Company in FDA record
Dako North America Inc.
FDA record creation date
Jul 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

To correct the kit's instructions for use as the storage time for gastric and gastroesophageal junction adenocarcinoma cut sections will be changed from six months to five months.

Source facts

Action
On February 2, 2018 the firm sent letters to the consignees with the following instructions; Description of the issue: The Field Correction has been initiated to correct the kit's Instructions for Use (I FU) as the storage time for Gastric and Gastroesophageal Junction (GEJ) Adenocarcinoma cut sections will be changed from six months to five months. This means that section 8.2: Cut Section Storage Recommendation will be changed to the following: 'To preserve antigenicity, tissue sections, once mounted on slides, should be held in the dark at 2-8 C (preferred), or at room temperature up to 25 C. Slide storage and handling conditions should not exceed 25 C at any point post-mounting to ensure tissue integrity and antigenicity. 8.2.1 NSCLC Cut Section Storage Recommendation Cut sections must be stained within 6 months when stored at 2-8 C (preferred), or at 25 C. 8.2.2 Gastric or Gastroesophageal Junction (GEJ) Adenocarcinoma Cut Section Storage Recommendation Cut sections must be stained within 5 months when stored at 2-8 C (preferred), or at 25 C." Note that only Gastric and Gastroesophageal Junction (GEJl Adenocarcinoma cut section storage time has been updated. Storage time for Non-Small Cell Lung Cancer (NSCLCl cut sections remains the ~ You may continue the use of PD-Ll IHC 22C3 pharmDx, Code No. SK006. However, we recommend that your laboratory follow the instructions noted to preseive the antigenicity, as this may decrease after five months of cut section storage. Potential patient risk: Based on our investigations, there is a remote possibility that staining Gastric and Gastroesophageal (GEJ) Adenocarcinoma cut sections may result in a weak staining or a false negative staining after five months of storage. However, there have been no reported incidents based on the previous storage time recommendation. Actions to be taken by the user: Within 10 calendar days, please take the following actions 1. Discard the IFU provided within the PD-Ll
Initiated
2018-02-02
Root cause
Process control
Terminated
2020-11-10
Distribution
AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, NH, NJ, NV, NY, OH, OK, PA, TX, UT, VA, WA, and WI
Recall number
Z-2425-2018
Recall status
Terminated
CDRH record ID
162750
FDA record created
2018-07-11
Recall event number
79602

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗