Open original FDA record ↗Recall record
- FDA identifier
- Z-2427-2018
- Company in FDA record
- Covidien LLC
- FDA record creation date
- Jul 12, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Devices were found with a sterile breach (puncture) in the pouch from a sort of 8,064 packaged devices.
Source facts
- Action
- Medtronic Regional sent a notification letter to customers via overnight delivery. the letter identified the affected product, problem and actions to be taken. The letter request that you quarantine and return any unused products of the affected item codes and production lots. Unused products from the affected item codes should be returned. If you have distributed Covidien LigaSure" Exact Dissector listed, please promptly forward the information from this letter to those recipients. All unused products from the affected item codes must be returned. The potentially affected product was shipped to your facility between November 2017 and February 2018.
- Initiated
- 2018-03-15
- Root cause
- Other
- Terminated
- 2019-12-11
- Distribution
- Worldwide Distribution in the countries of Australia, Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Spain, Swede, Switzerland and United Kingdom
- Recall number
- Z-2427-2018
- Recall status
- Terminated
- CDRH record ID
- 163231
- FDA record created
- 2018-07-12
- Recall event number
- 79735