Z-2429-2020: Oscar 2, Model 250 System

Recall record Z-2429-2020 for Suntech Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2429-2020
Company in FDA record
Suntech Medical, Inc.
FDA record creation date
Jun 25, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device display, when activated, can show an incorrect decimal point when the blood pressure reading is a three-digit number and the unit of measure is set for mmHg.

Source facts

Action
Suntech Medical Instruments, Inc. notified customers on about [DATE] via "URGENT: MEDICAL DEVICE RECALL" letter. Customers were informed that the device display, when activated, can show an incorrect decimal point when the blood pressure reading is a three-digit number and the unit of measure is set for mmHg. Suntech Medical is instructing physicians and nurses that they can still use the device and are to ignore the readings on the device display and to continue to focus on the Accuwin Pro data for diagnosis. Best practice is to disable the device display during unit configuration and to instruct the patient to ignore any data that may be visible on the display. If these steps are followed, Suntech states the device can continue to be used. The correction will include 2 options: 1. To download the service tool at https://www.suntechmed.com/oscar2-display-update-request; or 2. Customers should contact SunTech to confirm the configuration has been corrected at recallcoordinator@suntechmed.com and provide the contact person, company name, and serial numbers of corrected devices. Additionally, customers were requested to complete and return the response form via email to recallcoordinator@suntechmed.com, or they can be sent by mail to: Recall Coordinator, SunTech Medical, Inc., 507 Airport Blvd, suite 117, Morrisville, NC 27560-8200. Questions can be addressed to SunTech Medical at Email: recallcoordinator@suntechmed.com, Phone: +1-800-421-8626 or 919-654-6128
Initiated
2020-04-20
Root cause
Process change control
Terminated
2023-06-13
Distribution
Nationwide distribution to CA, CT, FL, IL, LA, MD, NC, NY, OH, PA, PR, SC, TN, TX, UT. Worldwide distribution to Australia, Austria, Belgium, Bermuda, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, Estonia, France, India, Ireland, Latvia, Lebanon, Morocco, Pakistan, Peru, Poland, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, United Kingdom, Vietnam.
Recall number
Z-2429-2020
Recall status
Terminated
CDRH record ID
181130
FDA record created
2020-06-25
Recall event number
85515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗