Open original FDA record ↗Recall record
- FDA identifier
- Z-2430-2018
- Company in FDA record
- Valeris Medical, LLC
- FDA record creation date
- Jul 13, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.
Source facts
- Action
- Valeris Medical notified customers on approximately 01/20/2018 via phone call, text, or email. A formal letter was mailed to customers on 02/09/2018. The letter identified the affected product, problem and actions to be taken. Customers were instructed to remove affected products from inventory to quarantine, affix replacement labels provided by Valeris over the current labeling, and notify any customers if product was further distributed to so the product can be quarantined and returned for labeling. Additionally, Valeris requested confirmation in writing that the field correction action has been completed.
- Initiated
- 2018-01-22
- Root cause
- Process control
- Terminated
- 2021-04-09
- Distribution
- US Nationwide Distribution in the states of California and Oklahoma.
- Recall number
- Z-2430-2018
- Recall status
- Terminated
- CDRH record ID
- 164354
- FDA record created
- 2018-07-13
- Recall event number
- 80069