Calibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.
Source facts
Action
Siemens Healthcare Diagnostics sent a letter notification sent to customers dated June 10, 2020. As one of the short-term alternatives for customers, where regulations permit, Siemens has evaluated the use of the lyophilized RANDOX Calibration Serum Level 3 Lot 1024UE. (CAL 3) for generating DBIL_2/DBil_2 and TBIL_2/TBil_2 results until a new lot of unaffected Siemens Chemistry Calibrator is available.
Initiated
2020-06-05
Root cause
Nonconforming Material/Component
Terminated
2021-11-18
Distribution
US Nationwide Distribution
Recall number
Z-2430-2020
Recall status
Terminated
CDRH record ID
181892
FDA record created
2020-06-26
Recall event number
85844