Z-2431-2023: Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-

Recall record Z-2431-2023 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2431-2023
Company in FDA record
Datascope Corp.
FDA classification date
Aug 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. Docking/Power Battery Failure User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra- Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. The Cardiosave device has two operating modes: Hybrid (docked within the hospital cart) and Rescue (transport). When in Hybrid mode the IABP may operate on both AC power and battery power; and when plugged in (AC power) can charge inserted batteries. When in Rescue (transport) mode, the IABP can operate on battery power, or AC Power with the use of the Transport Power Supply accessory (occupying a battery bay). However, unless the Cardiosave is powered off, the Transport Power Supply will not charge a battery inserted into the alternate battery bay unless the Transport Power Supply is connected directly to AC power. There are two methods where the IABP can be in transport/rescue mode without user intent: 1) The Cardiosave Hospital Cart's release latch has been released and the IABP console is not properly installed in the cart. 2) The IABP console is inserted into the hospital cart, but is not fully latched into the cart In both cases, information is provided to the user that the system is not in Hybrid mode via the Rescue Mode Icon on the bottom Cardiosave screen (top center of the screen). In addition, the battery in use message will be displayed in the informational message area even if the AC power cord is plugged in. Additionally, in both scenarios, the Cardiosave will not receive AC power and if the device is in operation, it will continue to operate on battery. When the Cardiosave is not receiving AC power, the batteries are unable to charge. If the AC Plug Icon is not present and/or the Rescue Icon is present, release the latch on the Hospital Cart located b
Initiated
2023-07-31
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-2431-2023
Recall status
Open, Classified
CDRH record ID
201169
FDA classified
2023-08-30
Recall event number
92501

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗