Z-2431-2024: Abbott Infinity 7, Implantable Pulse Generator, REF: 6662, SterileEO, Rx Only

Recall record Z-2431-2024 for Abbott Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2431-2024
Company in FDA record
Abbott Medical
FDA classification date
Jul 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Source facts

Action
On June 6, 2024, Abbott, issued a "Urgent Medical Device Correction" notification to affected consignees via mail. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. For information on when an ERI notification appears and how to make a determination with your patients on timing of IPG replacement, see the customer communication. 3. For recommended strategies for preventing lapse in therapy, see the customer communication. 4. Please inform your patients about this medical device correction letter. 5. Please return a completed Acknowledgement Form and maintain a record of this notice along with a copy of the completed Acknowledgement Form. 6. For questions about this issue or reported complaints, please contact your Abbott representative, or Abbott Technical Support at 1-800-727-7846. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2024-05-22
Root cause
Error in labeling
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-2431-2024
Recall status
Open, Classified
CDRH record ID
208348
FDA classified
2024-07-25
Recall event number
94796

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗