Skip to content

Public FDA record

Z-2432-2008: Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid

Recall record Z-2432-2008 for Remel, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2432-2008
Company in FDA record
Remel, Inc
FDA classification date
Sep 19, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Faint grey test lines for Cryptosporidium may be visible which may be interpreted as false positive

Source facts

Action
The recalling firm issued letters dated 6/19/08 flagged "IMPORTANT PRODUCT INFORMATION" which advised their customers of the issue with the faint grey test lines but did not request any action be taken, such as return or destruction of the affected lot numbers. The recalling firm issued a second letter flagged "IMPORTANT MEDICAL DEVICE CUSTOMER ADVISORY NOTICE" dated 8/11/08 and was issued via regular mail on 8/14/08. This letter asked users to review their inventory for the product lots indicated in the letter, discard all remaining units of this product lot, complete and return the Product Inventory Checklist regardless if they have affected product. A "PRODUCT INVENTORY CHECKLIST" was enclosed to report whether the firm had product to discard and the amount to be replaced.
Initiated
2008-08-11
Root cause
Process control
Terminated
2010-01-21
Distribution
Worldwide Distribution --- including USA and countries of Canada and U.K.
Recall number
Z-2432-2008
Recall status
Terminated
CDRH record ID
72732
FDA classified
2008-09-19
Recall event number
49090

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗