Open original FDA record ↗Recall record
- FDA identifier
- Z-2432-2020
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA record creation date
- Jun 26, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Source facts
- Action
- On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative
- Initiated
- 2020-05-27
- Root cause
- Software design
- Terminated
- 2021-01-12
- Distribution
- Foreign distribution only.
- Recall number
- Z-2432-2020
- Recall status
- Terminated
- CDRH record ID
- 181676
- FDA record created
- 2020-06-26
- Recall event number
- 85750