Open original FDA record ↗Recall record
- FDA identifier
- Z-2433-2021
- Company in FDA record
- Armstrong Medical Services Limited
- FDA classification date
- Sep 13, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Some devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthetised patient, which could cause a delay in treatment. Not all defective canisters will be identified by the pre-use test.
Source facts
- Action
- On August 5, 2021, the firm emailed Urgent Medical Device Correction letters to impacted customers. Users are requested to review the list of potentially affected devices and return the completed response form to Armstrong Medical or to an appointed distributor to receive replacement units. Where users have opted to temporarily retain their stock of potentially affected devices, those users are asked to ensure that the revised pre-use test in the letter is always completed prior to use. The letter included instructions to check that the ventilator functions correctly, to be performed in supplement to the <> test before each patient. Customers may contact the recalling firm at Quality@armstrongmedical.net.
- Initiated
- 2021-08-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-09-17
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Canada, Finland, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Portugal, United Kingdom.
- Recall number
- Z-2433-2021
- Recall status
- Terminated
- CDRH record ID
- 188724
- FDA classified
- 2021-09-13
- Recall event number
- 88462