Z-2433-2021: AMSORB PLUS PREFILLED G-CAN 1.0L; Carbon dioxide absorbent canister; Product codes AMAB3801 and AMAB3801G

Recall record Z-2433-2021 for Armstrong Medical Services Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2433-2021
Company in FDA record
Armstrong Medical Services Limited
FDA classification date
Sep 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Some devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthetised patient, which could cause a delay in treatment. Not all defective canisters will be identified by the pre-use test.

Source facts

Action
On August 5, 2021, the firm emailed Urgent Medical Device Correction letters to impacted customers. Users are requested to review the list of potentially affected devices and return the completed response form to Armstrong Medical or to an appointed distributor to receive replacement units. Where users have opted to temporarily retain their stock of potentially affected devices, those users are asked to ensure that the revised pre-use test in the letter is always completed prior to use. The letter included instructions to check that the ventilator functions correctly, to be performed in supplement to the <> test before each patient. Customers may contact the recalling firm at Quality@armstrongmedical.net.
Initiated
2021-08-05
Root cause
Nonconforming Material/Component
Terminated
2024-09-17
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Canada, Finland, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Portugal, United Kingdom.
Recall number
Z-2433-2021
Recall status
Terminated
CDRH record ID
188724
FDA classified
2021-09-13
Recall event number
88462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗