Z-2434-2008: Kwiktrak Gate System; the rail system used with the Maxi Sky 600 Ceiling Patient Lift; The Kwiktrak Gate

Recall record Z-2434-2008 for Arjo, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2434-2008
Company in FDA record
Arjo, Inc.
FDA classification date
Sep 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a possibility that the Kwiktrak Gate System could malfunction, resulting in the gate locks opening even though the corresponding tracks are not properly aligned.

Source facts

Action
Arjo Inc. USA sent Urgent Field Safety Notice letters dated 8/1/08 to the direct accounts on 8/1/08, informing them that there is a remote possibility that the gate system could malfunction. The result of this malfunction is that the gate locks could open even though the corresponding tracks are not properly aligned. This could result in the lift/cassette falling off the track system, possibly resulting in serious personal injuries and or property damage. The letters requested the customer to immediately implement special operating instructions in their facility to ensure great care is used when the gate system is operated. All personnel must be made aware that the tracks need to be properly aligned before using the gate. Illustrations of an unaligned gate system with an open lock and properly aligned tracks and gate system with open locks were included. Any questions the customers may have were directed to the Arjo Quality Department at 1-800-323-1245, ext. 6125. The accounts were informed that Arjo personnel will contact them in September 2008 to install a permanent add-on improvement to prevent the occurrence of a malfunction. The accounts were requested complete the enclosed response form indicating that they have read and understood the 8/1/08 letter, and fax it back to Arjo at 888-594-2756.
Initiated
2008-08-01
Root cause
Device Design
Terminated
2010-01-05
Distribution
Nationwide Distribution --- including states of Arizona, Colorado, Illinois, Indiana, Michigan, Minnesota, Missouri, Montana, New York, North Carolina, Ohio, Oklahoma, Texas, and Wisconsin.
Recall number
Z-2434-2008
Recall status
Terminated
CDRH record ID
72770
FDA classified
2008-09-20
Recall event number
49102

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗