Due to customer-experienced complaints on narrow assay range compared to prior lots and the user being out of assay range.

Source facts

Action
Between 10/14/2016 to 10/21/2016 a notification was emailed to the affected users informing them that Assay protocol had changed due to a Marketing decision to delete use of EDC, since it was a discontinued device (for over 10 years), creating a tighter assay range, but met QC specification. Due to customer-experienced complaints on narrow assay range compared to prior lots and the user being out of assay range, Technical Services began an MRA investigation. Investigation verified user complaint, and Lot was quarantined, and Rework action plans using EDC also were initiated for the device. This issue was found at a Users site, via use as system Quality Control material. The investigation focused on potential effects/risks, and cause. This issue was found at a Users site, via use as system Quality Control material. Customers reported a narrow assay range when compared to prior lots forcing some users out of the provided assay range. PUBLIC CONTACT: Title: Technical Services Manager Address: 1530 Lindbergh Drive, Beaumont, TX 77707 Telephone #: (409) 842-3714 x1177 Email: jhanka@helena.com
Initiated
2016-10-14
Root cause
Process control
Terminated
2023-10-12
Distribution
Domestic: FL, IL, MN, MT, NY, OH, TN, TX, VA, and WI OUS: Australia, Canada, Jordan, and South Korea
Recall number
Z-2434-2019
Recall status
Terminated
CDRH record ID
168437
FDA record created
2019-08-30
Recall event number
81383

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗