Z-2434-2021: Proteus 235- designed to produce and deliver a proton beam for the treatment of patients with localized t

Recall record Z-2434-2021 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2434-2021
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Sep 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment

Source facts

Action
IBA issued an Urgent Medical Device Correction notification users of the device, starting June 14, 2021. Letter states reason for recall, health risk and action to take: IBA would like to remind Proteus235 users to verify with X-ray images or other means, after the display of the popup message, that the patient positions (represented by the couch position, mainly the 3 translations and 3 rotations parameters) recorded in the OIS are correct Labelling change IBA will update the user manuals to: - Describe under which conditions the popup message appears and the subsequent risks, especially that the equipment positions may not be up to date in the OIS, - Remind the user to verify with X-ray images or other means, after the display of the popup message, that the patient positions (represented by the couch position, mainly the 3 translations and 3 rotations parameters) recorded in the OIS are correct, - Recommend the user to verify the OIS user manuals for measures preventing saving outdated patient positions in case of disconnection and to apply these measures before saving the positions in the OIS. The action will be completed for your site by May 31, 2022. By signing below, the customer representative confirms that this notice has been read, understood and communicated to the appropriate employees within the organization. Please transfer this notice to other organizations on which this action has an impact. Your IBA representative is able to provide you with additional information and/or guidelines if necessary. Please return the copy of the notice signed to IBA within 10 working days .
Initiated
2021-06-14
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, OK, VA, IL, NJ, TN, WA, LA, TX, MI and the countries of Germany, Italy, Russia, Sweden, India, Japan, France, England, Spain.
Recall number
Z-2434-2021
Recall status
Open, Classified
CDRH record ID
188701
FDA classified
2021-09-08
Recall event number
88449

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗