Reason stated in the source
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Source facts
- Action
- On June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.
- Initiated
- 2024-06-14
- Root cause
- Device Design
- Distribution
- Domestic distribution nationwide. International distribution to Canada.
- Recall number
- Z-2435-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208602
- FDA classified
- 2024-07-25
- Recall event number
- 94890