Open original FDA record ↗Recall record
- FDA identifier
- Z-2436-2010
- Company in FDA record
- Gyrus Acmi, Incorporated
- FDA classification date
- Sep 19, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Gyrus Medical, Inc. has issued a Urgent Medical Device Recall for the Gyrus ACMI product 004942-903 - Pleatman Sac Tissue Removal System, sold as lot 186540FC with an expiration date of 06/2015. As a result of a manufacturing error in the production of the trays used in this lot, it is possible that a crack could form that would compromise the package integrity, and possibly the sterility of the packaged device.
Source facts
- Action
- Gyrus Medical, Inc. of Maple Grove, MN issued an Urgent Medical Device Recall drawing attention to a possible package integrity compromise that could affect the product sterility. USA consignees were contacted via telephone on 8/3/10. The firm directed consignees to inspect their stock to confirm if any of the affected lot of product remains in their possession. Use of remaining affected product should cease immediately. Affected products should be return to Gyrus ACMI. For questions or concerns, consignees should contact the firm at (888) 524-7266.
- Initiated
- 2010-08-03
- Root cause
- Packaging process control
- Terminated
- 2012-08-23
- Distribution
- UT, PA, Canada
- Recall number
- Z-2436-2010
- Recall status
- Terminated
- CDRH record ID
- 94140
- FDA classified
- 2010-09-19
- Recall event number
- 56626