Z-2436-2020: Neo-Vent Infant T-Piece Resuscitation Circuit With Neonatal Mask; Circuit for use on compatible Infant Re

Recall record Z-2436-2020 for Westmed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2436-2020
Company in FDA record
Westmed, Inc.
FDA record creation date
Jun 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Source facts

Action
The firm sent a notification letter via email and it was issued on 1/15/2019. The following are actions given by the recalling firm to be taken by the customer/user: 1. If you are using the Neo-Vent Circuit with a 10 mm ID connection system (such as the F&P Neopuff System) the connection is compatible with all product lots. You may continue to use it with no issues. 2. If you are using the Neo-Vent Circuit with a 15 mm OD connection system (such as the GE Panda or Giraffe System), the connection may not be compatible with the list of LOT numbers. Quarantine the product and contact via the contact information provided. Reference numbers twill be provided to facilitate a return merchandise authorization (RMA). Contact info: Diana Upp at (520)294-7987 extension 1235 or Dean Iwasaki at extension 1132. Contact Information: Mail original signed documents to Regulatory Department at 5580 S. Nogales Hwy, Tucson AZ, 85706. Additionally, scan/email a copy of the documents to Diana Upp at dupp@westmedinc.com. Response is required; complete the Medical Device Recall Return Response Acknowledgement and Receipt Form Distributors are instructed to do the following: 1) Check stock and quarantine inventory. 2) Identify and notify customers that were shipped or may have been shipped product. 3) Attach a list of customers who received/may have received product. Customers of distributors to be notified.
Initiated
2019-01-15
Root cause
Device Design
Terminated
2021-02-05
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.
Recall number
Z-2436-2020
Recall status
Terminated
CDRH record ID
181266
FDA record created
2020-06-26
Recall event number
85576

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗