Open original FDA record ↗Recall record
- FDA identifier
- Z-2436-2021
- Company in FDA record
- Angiodynamics, Inc.
- FDA record creation date
- Sep 8, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up.
Source facts
- Action
- On July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees. Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices. In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere to the following Warning in the current product labeling (User Manual): WARNING: Do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "Ready".
- Initiated
- 2021-07-20
- Root cause
- Software design
- Terminated
- 2023-03-15
- Distribution
- Domestic distribution Nationwide. Foreign distribution worldwide.
- Recall number
- Z-2436-2021
- Recall status
- Terminated
- CDRH record ID
- 188680
- FDA record created
- 2021-09-08
- Recall event number
- 88444