Z-2436-2025: WATCHMAN TruSteer Access System, US, Material Number (UPN) M635TU90050;intended to provide vascular and t

Recall record Z-2436-2025 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2436-2025
Company in FDA record
Boston Scientific Corporation
FDA classification date
Aug 29, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

Source facts

Action
Boston Scientific issued an Important Medical Device Advisory notice to its consignees on 07/29/2025 via FedEx. The notice explained the problem with the device, procedures under which the likelihood of the problem occurs, and risk to the patient. The WATCHMAN Access Systems Instructions for Use (IFUs) and WATCHMAN physician training will be updated to emphasize instructions related to Access System air management. This update will strengthen the information provided to clinicians regarding the potential for air embolism during WATCHMAN procedures performed under conscious or deep sedation and provide potential mitigation strategies. No devices are being removed. Instructions: 1. Review IFU updates related to air embolism as detailed in Appendix 2. These updates will be added to the WATCHMAN Access Systems IFUs and WATCHMAN Physician Training. 2. Forward this letter to any other clinicians in your medical facility who perform WATCHMAN procedures and to any facilities where affected devices have been transferred, including hospitals or sites within your network. If you are a distributor, this notice must be forwarded to your customers to ensure notification of this Medical Device Advisory is carried out to the end-user level. 3. Complete and return the enclosed Acknowledgment Form. If additional assistance is required or more information regarding this communication, customer should contact your local Boston Scientific representative or Sr. Quality Systems Manager at 763-494-1133.
Initiated
2025-07-29
Root cause
Under Investigation by firm
Distribution
Worldwide.
Recall number
Z-2436-2025
Recall status
Open, Classified
CDRH record ID
215352
FDA classified
2025-08-29
Recall event number
97363

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗