Reason stated in the source
The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Source facts
- Action
- Consignees (Hospital Administrators; Hospital Risk Management Department; Mangers of Radiology/Cardiology; Radiologists/Cardiologists) were sent a "GE Urgent Medical Device Correction" letter dated July 2008. The letter described the affected products; the problem; and the user instructions, which state that the hospital should establish emergency workflow procedures to ensure patient safety should a component stop operating during a patient examination. Facilities should verify if their system has been impacted by checking if the catalog numbers listed are referenced on the PDB panel located in the equipment room. The letter also states the GE Healthcare Resolution Plan, includes correcting the condition, and contact Information.
- Initiated
- 2008-07-25
- Root cause
- Device Design
- Terminated
- 2011-12-10
- Distribution
- Nationwide Distribution --- including states of CA, FL, NY, AK, AL, AR, AZ, PA, FL ,CO, SC, CT, IA, DC, DE, GA, HI, IL , IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NJ, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and GUAM.
- Recall number
- Z-2437-2008
- Recall status
- Terminated
- CDRH record ID
- 72790
- FDA classified
- 2008-09-20
- Recall event number
- 49105