Open original FDA record ↗Recall record
- FDA identifier
- Z-2437-2020
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- Jun 26, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review
Source facts
- Action
- Siemens issued an Urgent Medical device Correction (UMDC ISW-20-01.A.US) and Urgent Field Safety Notification (UFSN ISW-20-01.A.OUS) to affected Customers via overnight mail on 5/15/20. The UMDC ISW 20-01.A.US and UFSN ISW-20-01. Letter states reason for recall, health risk and action to take: Manually Review QC results in CentraLink/Atellica DM daily; Review this letter with your Medical Director. If your laboratory may be affected by this issue based on the conditions listed above, you may consider the following steps to avoid this issue from occurring until Siemens updates your driver: Do not process QC while the interface communication is interrupted Use the Control lot number as the Control ID. Customers were requested to respond via the attached effectiveness check form to indicate receipt and understanding of the letter. Siemens will contact all affected customers to schedule a date to update the interface driver to version 1.5.1, which prevents the issue fromoccurring.
- Initiated
- 2020-05-15
- Root cause
- Software design
- Terminated
- 2021-05-18
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Recall number
- Z-2437-2020
- Recall status
- Terminated
- CDRH record ID
- 181660
- FDA record created
- 2020-06-26
- Recall event number
- 85737