Z-2438-2020: (1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data

Recall record Z-2438-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2438-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Jun 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

Source facts

Action
Siemens issued an Urgent Medical device Correction (UMDC ISW-20-01.A.US) and Urgent Field Safety Notification (UFSN ISW-20-01.A.OUS) to affected Customers via overnight mail on 5/15/20. The UMDC ISW 20-01.A.US and UFSN ISW-20-01. Letter states reason for recall, health risk and action to take: Manually Review QC results in CentraLink/Atellica DM daily; Review this letter with your Medical Director. If your laboratory may be affected by this issue based on the conditions listed above, you may consider the following steps to avoid this issue from occurring until Siemens updates your driver: Do not process QC while the interface communication is interrupted Use the Control lot number as the Control ID. Customers were requested to respond via the attached effectiveness check form to indicate receipt and understanding of the letter. Siemens will contact all affected customers to schedule a date to update the interface driver to version 1.5.1, which prevents the issue fromoccurring.
Initiated
2020-05-15
Root cause
Software design
Terminated
2021-05-18
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-2438-2020
Recall status
Terminated
CDRH record ID
181661
FDA record created
2020-06-26
Recall event number
85737

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗