In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.
Source facts
Action
PTW issues Urgent Safety Notice due to an error with firmware version 2.05 of the UNIDOS webline in some cases, especially with the high dose-per-pulse-rates typical for particle therapy, measuring errors can result.
Initiated
2015-05-27
Root cause
Software design
Terminated
2017-03-31
Distribution
Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.
Recall number
Z-2439-2015
Recall status
Terminated
CDRH record ID
137618
FDA classified
2015-08-21
Recall event number
71395