Skip to content

Public FDA record

Z-2440-2019: syngo.CT software version VA48A_SP5.5 in the following SOMATOM computed tomography models: SOMATOM Defini

Recall record Z-2440-2019 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2440-2019
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Aug 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may result in a delay in diagnosis and/or patient rescans

Source facts

Action
Siemens issued notification Urgent Medical Device Correction letter on 7/25/19 issued via electronic mail to consignees or by USPS certified mail if a consignee email address was not available. Letter states problem, health risk, action to take: The remote software update installation process will require approximately 30 minutes for completion. The process is completely automatic; make sure the system and power are stable before and during the process and do not switch off the system during the update process. Siemens highly recommends starting the installation when the scanner will not be in use or when the necessary time for the update to be completed can be scheduled. If your system is connected to Siemens Remote Connection, this update will be performed remotely. Following the corrective action, the cause will be eliminated and a recurrence of the identified issues is prevented. If you have any questions, please contact our service organization at 1-800-888-7436.
Initiated
2019-07-25
Root cause
Software design
Terminated
2022-12-29
Distribution
Nationwide Foreign:
Recall number
Z-2440-2019
Recall status
Terminated
CDRH record ID
175606
FDA record created
2019-08-30
Recall event number
83519

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗