Skip to content

Public FDA record

Z-2440-2020: Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 040338

Recall record Z-2440-2020 for LivaNova Deutschland GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2440-2020
Company in FDA record
LivaNova Deutschland GmbH
FDA record creation date
Jul 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Source facts

Action
The firm issued an Urgent Field Safety Notice, SCP Control panel (Subset) dated August 29, 2019 via USPS on August 29, 2019 to inform consignees that LivaNova is executing a voluntary medical device correction a limited subset of SCP control panels (references 60-02-15 & 60-02-50) due to an observed trend in reported complaints. The letter describes immediate action to be taken by customers. It states: LivaNova is initiating the proactive replacement of potentially affected angle encoder lots in the field in order to reduce the likelihood of the occurrence of the above described issue. Actions to be taken by the Customer/User Pending the replacement of affected angle encoder(s) would be completed, LivaNova is kindly requesting SCP system users to take the following actions, without any delay: 1. Affected SCP control panel(s) listed in Attachment 1 should be quarantined. Your LivaNova representative may support you in this activity (refer to contact reference person at the end of this letter). 2. In case of urgent medical need, affected devices could be used, provided the SCP system operating instructions would be strictly followed, especially safety instructions related to the pump operation. Users should: a. ensure to have a SCP emergency drive unit ready for use, b. be familiar with the mounting as well as the functioning of the emergency drive unit. As reported in the operating instructions, hand cranking remains an option in case it would become necessary to operate the SCP manually. 3. Please complete and return the Customer Response Form (Attachment 2) by e-mail to LivaNova.FSCA@livanova.com. For questions regarding this Field Safety Notice, please contact your LivaNova representative, or LivaNova Customer Quality organization at LivaNova.FSCA@livanova.com. A copy of this letter has been provided to the U.S. Food and Drug Administration (FD
Initiated
2019-08-26
Root cause
Device Design
Terminated
2021-04-22
Distribution
US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany
Recall number
Z-2440-2020
Recall status
Terminated
CDRH record ID
181121
FDA record created
2020-07-08
Recall event number
85505

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗