Skip to content

Public FDA record

Z-2440-2021: COPAN FLOQSwabs, STERILE.EO, for drawing clinical samples. CE0123, for the following reference numbers

Recall record Z-2440-2021 for Copan Italia. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2440-2021
Company in FDA record
Copan Italia
FDA classification date
Sep 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A sterility assurance level of 10-6 cannot be guaranteed due to intentional fraud and data integrity issues at the contract sterilizer, Steril Milano.

Source facts

Action
On July 26, 2021, Copan Italia issued a "Urgent: Medical Device Recall" notification to all affected consignee. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. This notice should be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. 2. Please identify, segregate, and discontinue use of all affected product included within ATTACHMENT #1_FLOQSwabs Recall product list . 3. Confirm you have read and understood the recall instructions provided in the letter in writing using the FORM #1_FLOQSwabs Recall Return Response , by email, to: Francisco Cuerva, Copan Diagnostics, Inc., 26055 Jefferson Avenue, Murrieta, CA US 92562, Phone: 800 216 4016 Ext. 204, InfoRecall@copanusa.net 4. Disposal must be performed according to local regulation and it does not require biohazard disposal precautions. Evidence or declaration of Product disposal must be returned in writing, using the FORM #2_FLOQSwabs Recall Disposal evidence , by email, to: Francisco Cuerva, Copan Diagnostics, Inc., 26055 Jefferson Avenue, Murrieta, CA US 92562, Phone: 800 216 4016 Ext. 204, InfoRecall@copanusa.net 5.If you are not in the position to identify, segregate or discard the Product you must advise in writing using the FORM #1_FLOQSwabs Recall Return Response , by email, to: Francisco Cuerva, Copan Diagnostics, Inc., 26055 Jefferson Avenue, Murrieta, CA US 92562, Phone: 800 216 4016 Ext. 204, InfoRecall@copanusa.net 6.This recall should be carried out to the end user level for all the affected Products that have been distributed including any products under private labeling, incorporated or utilized as kit components. 7. Any adverse events associated with recalled Product must be communicated in writing using the FORM #1_FLOQSwabs Recall Return Response , by email, to: Francisco Cuerva, Copan Diagnostics, Inc., 26055 Jefferson Avenue, Murri
Initiated
2021-07-09
Root cause
Vendor change control
Terminated
2024-01-25
Distribution
US nationwide distribution, including Puerto Rico.
Recall number
Z-2440-2021
Recall status
Terminated
CDRH record ID
188660
FDA classified
2021-09-09
Recall event number
88437

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗