Open original FDA record ↗Recall record
- FDA identifier
- Z-2441-2018
- Company in FDA record
- Zeiss, Carl Inc
- FDA record creation date
- Jul 19, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Source facts
- Action
- The Company will immediately notify by direct contact (when available) and/or express courier each first line purchaser and/or known user site with informing them of the risk, how to prevent it and the notice of action. The Company will dispatch factory trained service representatives to perform following actions: o Installation of Hot Fix # 14 (or newer) for ZEN 2.3 o Installation of Firmware 1 .209 (or newer) on nodes x l 5 and x l 9 (Main Controller 2)
- Initiated
- 2018-05-29
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2024-07-26
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2441-2018
- Recall status
- Terminated
- CDRH record ID
- 165394
- FDA record created
- 2018-07-19
- Recall event number
- 80418