Reason stated in the source
The graphics driver of the FDR Go PLUS might cause the appearance of a Blue Screen of Death (BSoD)
Source facts
- Action
- Fuji Film issued notifications via FedEx standard overnight mail on July 19, 2019 advising of the problem, health risk and action to take: During this field action, you can continue to use your current FDR Go PLUS by following these additional instructions:1. Set the monitor sleep mode to Off. Complete the Field Action Verification Form, and upon receipt FUJIFILM service personnel will contact you to schedule a visit to correct your device.FUJIFILM will correct your affected device(s) by applying a registry file to your current software version, at no cost to you.. Question contact: Field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email mailto:jeffrey.wan@fujifilm.com
- Initiated
- 2019-07-19
- Root cause
- Software design
- Terminated
- 2020-08-28
- Distribution
- Nationwide
- Recall number
- Z-2441-2019
- Recall status
- Terminated
- CDRH record ID
- 175614
- FDA record created
- 2019-08-30
- Recall event number
- 83524