Z-2441-2023: EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator

Recall record Z-2441-2023 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2441-2023
Company in FDA record
Boston Scientific Corporation
FDA classification date
Aug 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

Source facts

Action
Boston Scientific disseminated an Important Medical Device Information notice to the physician or healthcare professional consignees beginning on 07/11/2023 by hand delivery using their sales representatives, The notices explain the problem, risk, and recommend the following: Please distribute this letter to all HCPs who manage follow-ups of S-ICD patients. Perform the next scheduled device check in-person using a Model 3300 LATITUDE programmer with Model 3877 v1.04 software or Model 3200 EMBLEM programmer with Model 2877 v4.10 software to address any potential for this rare S-ICD behavior. Up-to-date product performance information and a device lookup tool are available within our Product Performance Resource Center at www.bostonscientific.com/ppr. If you have additional questions regarding this information or need to report an adverse event, please contact your Boston Scientific representative or Technical Services, and your local regulatory authority, as applicable. On 08/16/2023 Boston Scientific mailed the letter to EMBLEM S-ICD implanting and/or following physicians/clinics of record. On 09/15/2023, Boston Scientific began distributing a letter dated September 2023 (with the original communication enclosed), to new implanting and following physicians/clinics of record since the August 2023 communication.
Initiated
2023-07-11
Root cause
Software design
Distribution
Worldwide
Recall number
Z-2441-2023
Recall status
Open, Classified
CDRH record ID
202475
FDA classified
2023-08-22
Recall event number
92760

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗