Skip to content

Public FDA record

Z-2442-2021: MIVI Super 90 8F Guide Catheter, 95 cm, REF MIA-9095S-IDE, For Investigational Use Only; sterile, and Sup

Recall record Z-2442-2021 for Mivi Neuroscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2442-2021
Company in FDA record
Mivi Neuroscience Inc
FDA record creation date
Sep 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for nonsterility of product due to a possible defect in the pouch seal.

Source facts

Action
The recalling firm, MIVI Neuroscience, Inc., first notified clinical study sites of the recall by memo dated 7/23/2021 issued via email on 7/23/2021.The memo instructed the investigational study site to quarantine and return all of the affected catheters in their EVAQ Clinical Trial inventory. The site was informed the recalling firm is working to resolve the issue to provide replacement units to continue enrolling in the EVAQ Trial. They were instructed to not enroll any subjects until replacements are received. The recalling firm issued letters to their commercial customers dated 8/2/2021 on 8/2/2021 via email and certified mail. The customers were instructed to do the follow: Immediately examine your inventory and quarantine all products subject to this recall. All quarantined products should be returned to MIVI Neuroscience. 2. Please complete the Medical Device Recall Return Response attached. 3. Return the response form no later than 28 August 2021 to customerservice@mivineuro.com. 4. Return your unused inventory listed on the response form to the address listed below with the Return Authorization: RA66. 5. We will issue a credit for the devices returned. If you have any questions, contact VP Quality, Regulator and Clinical at (952) 944-3834 or email: customerservice@minineuro.com Monday through Friday 8:00am to 4:30pm, Central Time.
Initiated
2021-07-23
Root cause
Packaging
Terminated
2024-05-23
Distribution
US Distribution to states of: FL, MA, NJ, NY, and TN; and OUS (Foreign) distribution to countries of: France, Hungary, Slovenia, and Spain.
Recall number
Z-2442-2021
Recall status
Terminated
CDRH record ID
188894
FDA record created
2021-09-09
Recall event number
88408

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗