Open original FDA record ↗Recall record
- FDA identifier
- Z-2442-2023
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Aug 22, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Source facts
- Action
- Boston Scientific disseminated an Important Medical Device Information notice to the physician or healthcare professional consignees beginning on 07/11/2023 by hand delivery using their sales representatives, The notices explain the problem, risk, and recommend the following: Please distribute this letter to all HCPs who manage follow-ups of S-ICD patients. Perform the next scheduled device check in-person using a Model 3300 LATITUDE programmer with Model 3877 v1.04 software or Model 3200 EMBLEM programmer with Model 2877 v4.10 software to address any potential for this rare S-ICD behavior. Up-to-date product performance information and a device lookup tool are available within our Product Performance Resource Center at www.bostonscientific.com/ppr. If you have additional questions regarding this information or need to report an adverse event, please contact your Boston Scientific representative or Technical Services, and your local regulatory authority, as applicable. On 08/16/2023 Boston Scientific mailed the letter to EMBLEM S-ICD implanting and/or following physicians/clinics of record. On 09/15/2023, Boston Scientific began distributing a letter dated September 2023 (with the original communication enclosed), to new implanting and following physicians/clinics of record since the August 2023 communication.
- Initiated
- 2023-07-11
- Root cause
- Software design
- Distribution
- Worldwide
- Recall number
- Z-2442-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202476
- FDA classified
- 2023-08-22
- Recall event number
- 92760