Open original FDA record ↗Recall record
- FDA identifier
- Z-2443-2010
- Company in FDA record
- Osypka Medical Inc
- FDA classification date
- Sep 21, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a certain manufacturing period malfunction (i.e., not stimulate) when a type of 9 V batteries with higher than usual open circuit voltage (OCV) is used. Failure t
Source facts
- Action
- OSYPKA Medical sent "Urgent Medical Device Recall" letters dated June 30, 2010 and July 7, 2010 to the direct customers. The recall notice informed the customers of the Reason for Recall, the Potential Hazard, the Temporary Fix of the Problem, the Permanent Fix of the Problem and Recommended Actions to be taken by the customer. The customers were instructed to contact Osypka Medical via telephone (858) 454-0021 or e-mail to m.osypka@osypkamed.com to verify that the Model 3085 and/or PACE 203H is subject to this product removal, obtain a return-to-manufacturer-authorization (RMA) number ,and then return the affected devices to Osypka Medical. If you have any questions call (858) 454-0021 or e-mail to mail@osypkamed.com
- Initiated
- 2010-06-30
- Root cause
- Device Design
- Terminated
- 2012-08-10
- Distribution
- Nationwide Distribution: USA including states of CA & FL
- Recall number
- Z-2443-2010
- Recall status
- Terminated
- CDRH record ID
- 92700
- FDA classified
- 2010-09-21
- Recall event number
- 56121