Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli
Source facts
Action
Terumo CVS issued an URGENT MEDICAL DEVICE Recall letter dated September 10, 2012 to alert all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that a Terumo CVS representative will contact all affected US account.to schedule an appointment to replace the affected product. The affected product will be return to Terumo for destruction. Non-US Terumo Affiliates will be instructed to remove the devices from the field and provide a record of destruction. For questions call the Terumo CVS Customer Service at 1-800-521-2818
Initiated
2012-09-10
Root cause
Release of Material/Component prior to receiving test results
Terminated
2014-01-13
Distribution
Worldwide and US Nationwide Distribution
Recall number
Z-2443-2012
Recall status
Terminated
CDRH record ID
110196
FDA classified
2012-09-24
Recall event number
62313