Z-2443-2021: The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management a

Recall record Z-2443-2021 for Hero Health. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2443-2021
Company in FDA record
Hero Health
FDA record creation date
Sep 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to dispenser prompting for a missed dose that had already been dispensed

Source facts

Action
On 07/02/2021, the firm initially sent out an "URGENT NOTICE" via email notifying customers of an product update correction. Customers were informed that the firm had become aware that a recent firmware update (v2.4.12) that users have experienced an error, which caused improper recording of a dispensed dose as having not been dispensed. This error may prompt a user to dispense the same dose again, and the mobile app will indicate that the previously dispensed dose has not been taken yet. Customers are instructed to: -If the pill dispenser prompts you to re-dispense a scheduled dose which has already been dispensed today, select Skip. -Use extra caution when dispensing doses prior to the scheduled firmware update and verify that the pill dispenser has a working internet connection. This will ensure that the firmware update (v2.4.14) will be automatically applied on all potentially affected online pill dispensers (at approximately 2:00 AM on July 3, 2021) to correct the issue. -To manually update firmware immediately, follow these steps: 1) Ensure the Pill Dispenser is connected to WiFi (Check for the WiFi icon on the main screen). 2) Go to the Device menu, by pressing the Center button on the main screen. 3) Scroll down and select Device Settings . 4) Scroll down and select Firmware Update. 5) Select I m Ready. The update could take several minutes. It is recommended to run the firmware update at least 30 minutes before a scheduled dose. -If the pill dispenser is not connected to WiFi, the firmware update cannot be completed. In this event, contact the Recalling Firm's Member support for assistance before continuing to use the pill dispenser to ensure the update is applied as soon as possible For question or further assistance, contact the Recalling firm at 855-855-9962 or via email at hello@herohealth.com (24/7).
Initiated
2021-07-02
Root cause
Under Investigation by firm
Terminated
2023-11-02
Distribution
U.S. Nationwide distribution.
Recall number
Z-2443-2021
Recall status
Terminated
CDRH record ID
188593
FDA record created
2021-09-10
Recall event number
88389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗