Open original FDA record ↗Recall record
- FDA identifier
- Z-2445-2023
- Company in FDA record
- Invacare Corporation
- FDA classification date
- Aug 23, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user
Source facts
- Action
- Invacare initiated on June 15, 2023 via electronic communications Urgent Medical Device Field Correction to consignees. Letter states reason for recall, health risk and action to take: 1. Review your existing stock to locate, quarantine, and return the affected devices to Invacare. A list of serial numbers is attached to assist in this process. An example product label for is shown below with annotations. 2. For affected Dolomite Gloss rollators already delivered to a customer, please contact the customer immediately to inform them that the Dolomite Gloss rollator(s) that they purchased is/are part of a safety recall and that use of this product should be discontinued immediately. For convenience, we have attached a Customer Letter which you may use to communicate this information with your customer(s). 3. Invacare will work with you to provide a replacement rollator or credit for the affected rollator that is being returned. 4. Complete and return the enclosed Provider Acknowledgement Card by following the instructions on the card. If you do not have an acknowledgement card or require a new one, please contact Invacare at (877) 413-6008 (U.S. Customers), Monday Friday, 8 a.m. 5 p.m. EST, for assistance. 5. You are responsible for ensuring that this field correction is conducted with your customers. Customers who own affected units should be contacted and informed about this field correction. If you have any questions concerning these instructions, please call Invacare s field correction support line at (877) 413-6008 (U.S. Customers), Monday Friday, 8 a.m. 5 p.m. EST, for assistance.
- Initiated
- 2023-06-15
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2445-2023
- Recall status
- Open, Classified
- CDRH record ID
- 201935
- FDA classified
- 2023-08-23
- Recall event number
- 92645