Skip to content

Public FDA record

Z-2445-2023: DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model

Recall record Z-2445-2023 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2445-2023
Company in FDA record
Invacare Corporation
FDA classification date
Aug 23, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

Source facts

Action
Invacare initiated on June 15, 2023 via electronic communications Urgent Medical Device Field Correction to consignees. Letter states reason for recall, health risk and action to take: 1. Review your existing stock to locate, quarantine, and return the affected devices to Invacare. A list of serial numbers is attached to assist in this process. An example product label for is shown below with annotations. 2. For affected Dolomite Gloss rollators already delivered to a customer, please contact the customer immediately to inform them that the Dolomite Gloss rollator(s) that they purchased is/are part of a safety recall and that use of this product should be discontinued immediately. For convenience, we have attached a Customer Letter which you may use to communicate this information with your customer(s). 3. Invacare will work with you to provide a replacement rollator or credit for the affected rollator that is being returned. 4. Complete and return the enclosed Provider Acknowledgement Card by following the instructions on the card. If you do not have an acknowledgement card or require a new one, please contact Invacare at (877) 413-6008 (U.S. Customers), Monday Friday, 8 a.m. 5 p.m. EST, for assistance. 5. You are responsible for ensuring that this field correction is conducted with your customers. Customers who own affected units should be contacted and informed about this field correction. If you have any questions concerning these instructions, please call Invacare s field correction support line at (877) 413-6008 (U.S. Customers), Monday Friday, 8 a.m. 5 p.m. EST, for assistance.
Initiated
2023-06-15
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-2445-2023
Recall status
Open, Classified
CDRH record ID
201935
FDA classified
2023-08-23
Recall event number
92645

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗