Skip to content

Public FDA record

Z-2445-2024: LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

Recall record Z-2445-2024 for LINK BIO CORP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2445-2024
Company in FDA record
LINK BIO CORP
FDA classification date
Jul 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.

Source facts

Action
On June 27, 2024 URGENT: MEDICAL DEVICE RECALL letters were sent via email to distributors. Actions to be taken by the Customer/User: Please check your inventory for this item and lot combination. If the markings are incorrect, as indicated in Section 5 below, please return the affected product to LinkBio. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return any affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned, should any affected product be identified. Please return any affected product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2024-06-27
Root cause
Process change control
Distribution
US Nationwide distribution in the states of FL, GA, KS, OH, WI.
Recall number
Z-2445-2024
Recall status
Open, Classified
CDRH record ID
208826
FDA classified
2024-07-25
Recall event number
94926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗