Skip to content

Public FDA record

Z-2446-2008: Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphi

Recall record Z-2446-2008 for Smith & Nephew Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2446-2008
Company in FDA record
Smith & Nephew Inc
FDA classification date
Sep 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping.

Source facts

Action
Prior to the dissemination of the official notices, the firm notified personnel at our field offices of the BD syringe recall and advised that a subsequent recall of our Caption devices would follow. High volume users were contacted via phone on July 30th and the 31st and personnel at all affected field offices were notified via email on July 30, 2008 with an Urgent Medical Device Recall Letter. The firm requedsted that usere immediately cease distribution, examin the stock on hand and the stock of the accounts that have been suppplied and remove any affected ecices from all inventory. A verification notification was included with letter
Initiated
2008-08-05
Root cause
Other
Terminated
2010-05-28
Distribution
Nationwide Distribution
Recall number
Z-2446-2008
Recall status
Terminated
CDRH record ID
72817
FDA classified
2008-09-20
Recall event number
49129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗