Skip to content

Public FDA record

Z-2447-2019: Flexible Intubation Fiberscope, Model # 11301AB1

Recall record Z-2447-2019 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2447-2019
Company in FDA record
Karl Storz Endoscopy
FDA record creation date
Aug 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The action is being initiated due to intubation scopes improperly manufactured with a double port luer instead of a single port luer.

Source facts

Action
The firm mailed a "safety alert" on April 24, 2019 via FedEx. 2nd day delivery. The safety alert advised customers to do the following: 1. If you use the following automated processes, the high level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed: Reprocessing Method Impact, Medivators AER * Patient Safety may be Compromised, STERIS System 1E (SS1E)** Patient Safety may be Compromised The above connectors can only connect onto one port, therefore leaving the other port open. Depending on the reprocessing system, the cycle could result in an error or in successful completion of the cycle. " If an error code is displayed, the user would know the cycle is not complete and the scope has not been successfully reprocessed. " If the cycle is completed without an error code, the high-level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed. Improper reprocessing may result in incomplete decontamination, which may result in a cross-contamination patient infection. If you use Medivators AER or STERIS SS1E to reprocess this scope, please be aware of the possible risks and take the necessary actions to address any patient safety concerns. 2. Please complete the attached Action Requested form and follow the instructions outlined on the form. 3. you have any questions please contact the following person: This Recall Action, Emily Cuadros, Complaint Handling , 424-218-8289 Reprocessing, Shaun McGinley, Director of Reprocessing, 424-218-8247 Patient Safety, Dr. Dennis Fowler, Chief Medical Officer, 424-218-8603
Initiated
2019-04-24
Root cause
Process control
Terminated
2020-12-04
Distribution
US: VA, MO, CA, OH, NM, NY, CO, LA, ME, OUS: Canada
Recall number
Z-2447-2019
Recall status
Terminated
CDRH record ID
175600
FDA record created
2019-08-31
Recall event number
83419

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗