Z-2447-2023: The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASON

Recall record Z-2447-2023 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2447-2023
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Aug 23, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure.

Source facts

Action
Olympus issued Urgent Medical Device Field Corrective Action letter on July 12,2023. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the referenced devices and identify any device with the model names specified above. Please check all areas of the hospital and include a copy of this letter and addendum with any of these devices remaining in inventory. 2. Carefully read the content of this Medical Device Field Corrective Action as well as the attached Addendum for valve inspection instructions and updated MAJ-1444 reprocessing instructions: Disinfect the MAJ-1444 in the OER by placing the MAJ-1444 individually in the washing case of the endoscope reprocessor, with no other accessories. Note: Steam sterilization (autoclaving) or manual high-level disinfection remain available options per the existing instruction for use. 3. Ensure all personnel are informed on valve inspection requirements and to cease use and dispose of any valves suspected of having an abnormality. 4. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0428 5. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Actions to be taken by the company: Olympus is updating the MAJ-1444 Instructions for Use to reflect the information found in the attached addendum. MAJ-1444 valves will be shipped with the Addendum until the Instructions for Use are updated. Olympus requests that you report complaints, including any injuries associated with the use of MAJ-1444, to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). For additional support concerning this matter, please contact Tara Safi, Field Corrective Actions Administrator, at Tara.Safi@Olympus.com.
Initiated
2023-07-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2447-2023
Recall status
Open, Classified
CDRH record ID
202296
FDA classified
2023-08-23
Recall event number
92728

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗