Z-2447-2025: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Cata

Recall record Z-2447-2025 for Dexcom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2447-2025
Company in FDA record
Dexcom, Inc.
FDA classification date
Sep 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Source facts

Action
On 07/24/2025, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" via in-app messaging informing users that a software bug was identified with users using the following app version: -Dexcom ONE+ iOS/Android App version 1.4 and all earlier versions, -Dexcom G7 iOS/G7 Watch iOS/Android App versions 2.8 and earlier The software bug can cause an issue where, when a transmitter error occurs, the app terminates the sensor session and prompts the user to "start a new sensor" without providing a "sensor failed" alert. Customer are instructed that an upgrade from app version version the affected software versions is mandatory and users will not be able to to use the affected app versions after August 20, 2025. To update their app now to continue to use the app: 1. Tap Update App to go to the app store. 2. Install the latest Dexcom ONE+ app version or Dexcom G7 app version 3. Open the Dexcom ONE+ app or the G7 app If assistance is needed, contact Technical Support U.S.: 1-844-478-1600 Global: Find local Technical Support contact information at Dexcom.com
Initiated
2025-07-24
Root cause
Software design
Distribution
SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW12300 G7 iOS CGM App Worldwide Distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW14245 Dexcom ONE+ Android Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Israel, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden SW14244 D1G7 iOS Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden
Recall number
Z-2447-2025
Recall status
Open, Classified
CDRH record ID
215006
FDA classified
2025-09-04
Recall event number
97262

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗