Z-2448-2021: Colonoscope Family # 1-Pentax Video Colonoscopes (EC Family) with One Instrument Channel and a Water Jet

Recall record Z-2448-2021 for Pentax of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2448-2021
Company in FDA record
Pentax of America Inc
FDA record creation date
Sep 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updated Reprocessing Instruction For Use (rIFU): Impacted PENTAX Medical Video Upper GI Scopes (EG) and Video Colonoscopes (EC) families

Source facts

Action
Pentax issued Urgent Medical Device Correction letter via UPS mailing on 7/14/21. Letter states reason for recall, health risk and action to take (USB drive along with the enclosed instruction sheet enclosed). Letter states: PENTAX separated the EC and EG rIFUs into 5 separate rIFUs . The grouping will allow the following for reprocessing personnel: " Eliminate the need to identify optional configurations of instrument channels and water jet channels " Simplify the instructions and graphics in the individual rIFUs Please replace any previous versions of rIFUs with the most current revision. Please note that this documentation set is current as of this date. For future updates, please refer to the PENTAX online IFU library at https://ifu.pentaxmedical.com. Customer Instructions: Please complete the enclosed Field Correction Response Form upon receipt of this package, and email to PENTAX at customeradvisories@pentaxmedical.com. Questions regarding this action, please feel free to contact us at: o Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) o Fax: 201-799-4063 (alternate 201-391-4189) o Email: customeradvisories@pentaxmedical.com
Initiated
2021-07-14
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-2448-2021
Recall status
Open, Classified
CDRH record ID
188668
FDA record created
2021-09-10
Recall event number
88440

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗