Open original FDA record ↗Recall record
- FDA identifier
- Z-2448-2023
- Company in FDA record
- Randox Laboratories Ltd.
- FDA classification date
- Aug 23, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
Source facts
- Action
- Randox Labs Ltd (Manufacturer) extended recall to the US distributor via Medical Device Correction Letter dated July 19, 2023. The distributor initial contact with customer and follow up contacts via email and telephone. Letter states reason for recall, health risk and action to take: " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns please contact Rand ox Technical Services. Any customers using batches 590858 and 590859 for Ferritin, please contact technical.services@randox.com
- Initiated
- 2023-07-19
- Root cause
- Under Investigation by firm
- Distribution
- CA, FL, IL, ME, MI, NC, OH, PA,NJ,
- Recall number
- Z-2448-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202567
- FDA classified
- 2023-08-23
- Recall event number
- 92781