Z-2448-2026: HarmonyAIR A-Series Surgical Lighting System

Recall record Z-2448-2026 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2448-2026
Company in FDA record
Steris Corporation
FDA classification date
Jun 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 5/1/26 was sent to customers. STERIS Action All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their lighting system(s) to perform the correction as soon as possible. For the paint issue, each affected lighting system will undergo a full visual inspection for any signs of paint delamination; as needed a service kit will be installed to prevent paint chipping from occurring or if deemed necessary replaced. For the button issue, each affected lighting system will undergo inspection for any signs of potential separation from the light head and handle. Buttons failing this inspection will be replaced. User Action Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the Operator Manuals. Should any issues be observed prior to the correction being completed, users should submit a complaint by contacting STERIS Customer Service at 1-800-548-4873. Users are also reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact your local Account Manager.
Initiated
2026-05-15
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, India, New Zealand, Panama.
Recall number
Z-2448-2026
Recall status
Open, Classified
CDRH record ID
219748
FDA classified
2026-06-15
Recall event number
98767

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗