Abbott Vascular is recalling the Armada 35 and Armada 35 LL PTA Catheters because they have discovered that some devices may exhibit difficulty inflating and/or deflating.
Source facts
Action
An "Urgent Device Recall" letter dated 8/20/2012 was sent to all customers who purchased the Armada 35 and Armada 35 LL PTA Catheters. The letter informed the customers of the problem identified and the action to be taken. A list of lot numbers and an effectiveness check form was sent to customers with the customer notification letter. Customers were instructed to call Abbott Vascular Customer Service at (800) 227-9902.
Initiated
2012-08-20
Root cause
Process design
Terminated
2013-07-15
Distribution
Nationwide(US) Distribution and Puerto Rico.
Recall number
Z-2450-2012
Recall status
Terminated
CDRH record ID
112530
FDA classified
2012-09-25
Recall event number
63092