Skip to content

Public FDA record

Z-2451-2024: Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318

Recall record Z-2451-2024 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2451-2024
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jul 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.

Source facts

Action
On June 27, 2024 URGENT MEDICAL DEVICE REMOVAL - IMMEDIATE ACTION REQUIRED letters were sent to consignees. Actions to be taken: This removal affects only the products listed in Attachment 1 (Affected Products). No other material numbers or lots are impacted by this removal. Further distribution or use of any remaining product affected by this removal should cease immediately. For Capio SLIM devices that were successfully used in a procedure, no action is needed. Affected worldwide regulatory authorities are being notified of this removal as required. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. Share this notice with any health care professional within your organization and with any organization where the affected devices may have been transferred. If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. Patient safety is our highest priority. As such, we are committed to transparent communication to ensure that you have timely, relevant information for managing your patients. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative. Health care professionals and consumers may report serious adverse events (side effects) or product quality problems with the use of this product to Boston Scientific by calling 1-800-811-3211 and to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, fax, or phone. Online: www.fda.gov/MedWatch/report.htm Regular Mail: use postage-paid FDA form 3500 available at www.fda.gov/MedWatch/getforms.htm and mail to MedWatch, 5600 Fishers Lane, Rockvi
Initiated
2024-06-27
Root cause
Process change control
Distribution
Domestic: Nationwide Distribution. International: Pending
Recall number
Z-2451-2024
Recall status
Open, Classified
CDRH record ID
208819
FDA classified
2024-07-30
Recall event number
94921

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗