Skip to content

Public FDA record

Z-2451-2025: Akreos, SKU: AO60P0300. Akreos intraocular lenses

Recall record Z-2451-2025 for Bausch & Lomb Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2451-2025
Company in FDA record
Bausch & Lomb Surgical, Inc.
FDA classification date
Aug 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need for optical or surgical correction.

Source facts

Action
6/6/2025, recall notices were emailed to customers who were asked to do the following: 1) Check your inventory, quarantine, and do not use the IOL from this impacted lot. 2) Provide this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (as appropriate). Also, please transfer this notice to other organizations on which this action has an impact (as appropriate). Please maintain awareness of this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. 3) Complete and return the acknowledgement form via email to BLSurgCustSvc@Bausch.com For questions regarding this notice, please call firm at 1-800-338-2020, Option 9.
Initiated
2025-06-06
Root cause
Process control
Distribution
US: AZ, CA
Recall number
Z-2451-2025
Recall status
Open, Classified
CDRH record ID
214931
FDA classified
2025-08-26
Recall event number
97232

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗