Open original FDA record ↗Recall record
- FDA identifier
- Z-2451-2025
- Company in FDA record
- Bausch & Lomb Surgical, Inc.
- FDA classification date
- Aug 26, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need for optical or surgical correction.
Source facts
- Action
- 6/6/2025, recall notices were emailed to customers who were asked to do the following: 1) Check your inventory, quarantine, and do not use the IOL from this impacted lot. 2) Provide this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (as appropriate). Also, please transfer this notice to other organizations on which this action has an impact (as appropriate). Please maintain awareness of this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. 3) Complete and return the acknowledgement form via email to BLSurgCustSvc@Bausch.com For questions regarding this notice, please call firm at 1-800-338-2020, Option 9.
- Initiated
- 2025-06-06
- Root cause
- Process control
- Distribution
- US: AZ, CA
- Recall number
- Z-2451-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214931
- FDA classified
- 2025-08-26
- Recall event number
- 97232