Z-2451-2026: Medtronic O-arm O2 Imaging System. Model Number: BI70002000.

Recall record Z-2451-2026 for Medtronic Navigation, Inc.-Boxborough. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2451-2026
Company in FDA record
Medtronic Navigation, Inc.-Boxborough
FDA classification date
Jun 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.

Source facts

Action
Medtronic notified consignees on about 05/19/2026 via "URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION" letter. Consignees were informed that affected systems may continue to be used prior to service by Medtronic to update the firmware. The issue is intermittent and may not present, however if it does it can be resolved by power cycling (rebooting) the system. A Medtronic Field Service Engineer will contact each facility to schedule time to update the firmware on each system to permanently resolve this issue. Consignees were instructed to complete and return the provided Customer Confirmation Form, provide the notification to all applicable personnel within the organization or to customers if product has been transferred, and to maintain a copy of the notice with their records.
Initiated
2026-05-19
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Libya, Malaysia, Mexico, Morocco, Nepal, Netherlands, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-2451-2026
Recall status
Open, Classified
CDRH record ID
220683
FDA classified
2026-06-15
Recall event number
99041

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗