Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.
Source facts
Action
The firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.
Initiated
2008-08-05
Root cause
Nonconforming Material/Component
Terminated
2008-11-07
Distribution
Nationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.
Recall number
Z-2452-2008
Recall status
Terminated
CDRH record ID
72845
FDA classified
2008-09-21
Recall event number
49139